Hydrogen Sensor

The hydrogen sensor continuously detects the hydrogen proportion of the GC oven air. LEDs signal an increasing hydrogen concentration. If this concentration exceeds the limits, optical and acoustic alarms and the switch to inert gas will eliminate any risk of explosion. The Hydrogen Sensor can fully be integrated in Agilent 6890/7890/8890/8860 GCs and Shimadzu GCs 2010 (LEDs in front panel). For all other GC models, the sensor is available as external device.

Wasserstoff-Sensor für GC
HS 1000 10 00
H2 Sensor Neu

Function

  • Sensor configuration: Sensor head / control unit with LEDs, optical and acoustic alarm / gas switching module
  • Hydrogen concentration exceeds lower limit (10% of the lower explosion limit, LEL): first red LED lights up, audio warning
  • Hydrogen concentration exceeds upper limit (25% of the LEL resp. 1% by vol. H2): alarm signal sounds, automatic switching to inert gas
article order no.
Integrated H2-Sensor with Optical/Acoustic Alarm
(Sensor-Kit and Carrier Gas Change-Over Switch not included)
HS 2000 10 00
Agilent 8890/8860 GC Carrier Gas Change-Over Switch HS 2000 10 00 OPT026
Agilent 8890/8860 GC Sensor-Kit HS 2000 10 00 OPT016
Agilent 7890 GC Sensor-Kit HS 2000 10 00 OPT014
Agilent 7890 GC Carrier Gas Change-Over Switch HS 2000 10 00 OPT024
Agilent 6890 GC Sensor-Kit HS 2000 10 00 OPT012
Agilent 6890 GC Carrier Gas Change-Over Switch HS 2000 10 00 OPT022
Shimadzu GC 2010 Sensor-Kit (Standard pos. 10) HS 2000 10 00 OPT010
Shimadzu Carrier Gas Change-Over Switch HS 2000 10 00 OPT020
 
External H2-Sensor incl. Carrier Gas Switch (Sensor-Kit not included) HS 1000 10 00
Agilent 8890 GC Sensor-Kit HS 1000 10 00 OPT026
Agilent 7890 GC Sensor-Kit HS 1000 10 00 OPT014
Agilent 7820 GC Sensor-Kit HS 1000 10 00 OPT015
Agilent 6890 GC Sensor-Kit HS 1000 10 00 OPT012
Agilent 6850 GC Sensor-Kit HS 1000 10 00 OPT013
Shimadzu GC 2014 Sensor-Kit HS 1000 10 00 OPT030
Shimadzu GC 2010 Sensor-Kit HS 1000 10 00 OPT010
Perkin Elmer GC Sensor-Kit HS 1000 10 00 OPT017
Trace GC Sensor-Kit HS 1000 10 00 OPT016
Varian/Bruker GC Sensor-Kit HS 1000 10 00 OPT022

pdfHydrogen-Sensor.pdf